Olympus OER Pro Error Codes
Equipment support / Error reference
Olympus OER-Pro Error Codes & Troubleshooting Guide
Find an error code, review its possible cause, and see what to check first. This reference covers 39 OER-Pro codes, with expandable guidance for routine troubleshooting checks.
Search Olympus OER-Pro error codes
Search descriptions, causes and checks. Search and category filters work together. Some codes appear in more than one category.
39 of 39 codes shown
| Code | What it means | First checks & next action |
|---|---|---|
| E00Electrical / System | Cycle/process interrupted | Allow automatic error processing to finish. Follow the required purge/rinse procedure before proceeding.
Possible cause: STOP/RESET was pressed during operation. |
| E01Water / Drain | Basin takes too long to fill with water | Fully open the water supply. Check the supply hose, water filter and inlet mesh filter.E01: cause & checksE01 — Basin takes too long to fill with waterPossible cause: Water supply restricted, faucet not fully open, filter restriction or supply leak. First checks: Fully open the water supply. Check the supply hose, water filter and inlet mesh filter. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E02Water / Drain | Cleaning fluid will not drain | Check drain-hose installation and clean the basin drain mesh filter.E02: cause & checksE02 — Cleaning fluid will not drainPossible cause: Drain hose problem or clogged basin drain filter. First checks: Check drain-hose installation and clean the basin drain mesh filter. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E04Water / Drain | Cleaning-fluid level drops unexpectedly | Service recommended Service recommended. Possible cause: Internal machine problem. |
| E05Water / Drain | Cleaning-fluid level too high | Stop water supply if necessary and have the unit serviced.
Possible cause: Fluid-level sensing problem. |
| E06Sensors | Middle fluid-level sensor error | Clean the fluid-level sensor according to the manufacturer's maintenance instructions.E06: cause & checksE06 — Middle fluid-level sensor errorPossible cause: Residue on fluid-level sensor. First checks: Clean the fluid-level sensor according to the manufacturer's maintenance instructions. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E07Sensors | Lower fluid-level sensor error | Clean the fluid-level sensor.E07: cause & checksE07 — Lower fluid-level sensor errorPossible cause: Residue on fluid-level sensor. First checks: Clean the fluid-level sensor. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E11Water / Drain | Too much disinfectant in disinfectant tank | Check drain hose and basin drain filter. Verify disinfectant concentration before continued use.E11: cause & checksE11 — Too much disinfectant in disinfectant tankPossible cause: Drain or disinfectant-return problem. First checks: Check drain hose and basin drain filter. Verify disinfectant concentration before continued use. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E12Disinfectant | Too little disinfectant in tank | Clean circulation-port filters. Replace the disinfectant solution as required rather than topping off used solution.E12: cause & checksE12 — Too little disinfectant in tankPossible cause: Insufficient disinfectant / circulation restriction. First checks: Clean circulation-port filters. Replace the disinfectant solution as required rather than topping off used solution. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E13Water / Drain | Basin fills with disinfectant unusually quickly | Check drain hose and basin drain filter; remove residual solution and prepare fresh disinfectant as directed.E13: cause & checksE13 — Basin fills with disinfectant unusually quicklyPossible cause: Fluid remained in basin from an incomplete drain. First checks: Check drain hose and basin drain filter; remove residual solution and prepare fresh disinfectant as directed. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E14Water / Drain | Disinfectant cannot be returned to tank | Check gas-filter ventilation openings and clean the drain- and circulation-port mesh filters.E14: cause & checksE14 — Disinfectant cannot be returned to tankPossible cause: Drain/circulation filters restricted or gas-filter ventilation blocked. First checks: Check gas-filter ventilation openings and clean the drain- and circulation-port mesh filters. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E15Disinfectant | Fluid-level sensor error | Clean the fluid-level sensor. If the error returns, service may be required.
Possible cause: Residue on sensor or internal fault. |
| E16Disinfectant | Basin takes too long to fill with disinfectant | Service recommended Service recommended. Possible cause: Internal machine fault. |
| E17Disinfectant | Disinfectant will not heat correctly | Service recommended Service recommended. Possible cause: Internal heating-system fault. |
| E18Disinfectant | Temperature sensor error | Service recommended Service recommended. Possible cause: Internal temperature-sensing fault. |
| E21Flow / Pressure | Low air-purge pressure / air not flowing through scope channels | Verify air-filter installation. Replace the air filter if clogged.E21: cause & checksE21 — Low air-purge pressure / air not flowing through scope channelsPossible cause: Air filter installed incorrectly or restricted. First checks: Verify air-filter installation. Replace the air filter if clogged. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E22Flow / Pressure | Insufficient fluid pressure | Clean the circulation-port mesh filters.E22: cause & checksE22 — Insufficient fluid pressurePossible cause: Circulation-port filters restricted. First checks: Clean the circulation-port mesh filters. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E23Flow / Pressure | Excessive fluid pressure | Service recommended Service recommended. Possible cause: Internal equipment problem. |
| E31Lid / Drawer | Lid open | Close the lid firmly and restart. If it occurs after the cycle begins, service may be required.E31: cause & checksE31 — Lid openPossible cause: Lid not fully closed or lid-detection problem. First checks: Close the lid firmly and restart. If it occurs after the cycle begins, service may be required. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E41Electrical / System | Power lost and restored during cycle | Inspect the power connection and cord. Because the cycle was interrupted, follow the manufacturer's reprocessing instructions before reuse.
Possible cause: Power interruption, loose connection or damaged cord. |
| E51Water / Drain | Water detected leaking inside machine | Remove from service Close the water supply and remove the unit from service pending repair. Possible cause: Internal leak. |
| E61Electrical / System | Ultrasonic cleaning not operating | Service recommended Service recommended. Possible cause: Ultrasonic-system fault. |
| E71Disinfectant | Disinfectant-bottle drawer sensor error | Service recommended Service recommended. Possible cause: Internal drawer-sensing problem. |
| E72Water / Drain | Disinfectant tank will not drain | If solution remains in the basin, follow the E13 drain procedure. Otherwise, service is required.
Possible cause: Internal fault or retained solution. |
| E73Water / Drain | Water-supply problem while preparing disinfectant | Fully open water supply; inspect hose and inlet mesh; replace water filter if needed.E73: cause & checksE73 — Water-supply problem while preparing disinfectantPossible cause: Restricted water supply, leak, clogged inlet filter or water filter. First checks: Fully open water supply; inspect hose and inlet mesh; replace water filter if needed. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E74Disinfectant | Disinfectant from cassette bottles isn't reaching the unit | Verify the supply situation; persistent E74 generally requires service.
Possible cause: Cassette-supply or internal system problem. |
| E75Water / Drain | Water-supply problem while preparing disinfectant | Check faucet, water-supply hose, inlet mesh and water filter.E75: cause & checksE75 — Water-supply problem while preparing disinfectantPossible cause: Restricted or temporarily insufficient water supply. First checks: Check faucet, water-supply hose, inlet mesh and water filter. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E76Disinfectant | Disinfectant-tank fluid-level sensor error | Service recommended Service recommended. Possible cause: Internal sensor/system fault. |
| E77Disinfectant | Disinfectant bottle drawer not fully positioned | Push the drawer fully into its closed position.E77: cause & checksE77 — Disinfectant bottle drawer not fully positionedPossible cause: Drawer not completely closed. First checks: Push the drawer fully into its closed position. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E78Disinfectant | Cassette-bottle problem / bottles empty | Verify cassette installation and replace empty cassette bottles.E78: cause & checksE78 — Cassette-bottle problem / bottles emptyPossible cause: Bottles empty, improperly installed or not detected. First checks: Verify cassette installation and replace empty cassette bottles. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E81Electrical / System | Cycle cannot be properly controlled | Service recommended Service recommended. Possible cause: Internal machine fault. |
| E82Electrical / System | Internal machine irregularity | Service recommended Service recommended. Possible cause: Electrical/internal fault. |
| E83Electrical / System | Cycle cannot be properly controlled | Service recommended Service recommended. Possible cause: Internal electrical fault. |
| E84Electrical / System | RFID reader malfunction | Service recommended Service recommended. Possible cause: RFID/electrical fault. |
| E91RFID / ID | ID read error | Clear/reset the reading and scan the required ID again.E91: cause & checksE91 — ID read errorPossible cause: ID not detected or multiple IDs detected. First checks: Clear/reset the reading and scan the required ID again. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E92Flow / Pressure | Leak test exceeds allowed time | Verify the leak-test setup and perform the test again.
Possible cause: Leak-test process exceeds approximately 10 minutes. |
| E93Chemicals | Alcohol unavailable or not feeding | Check alcohol level and supply tubing/connections. Refill as needed; persistent errors require further diagnosis.E93: cause & checksE93 — Alcohol unavailable or not feedingPossible cause: Alcohol empty, supply restriction, sensor/pump/piping issue. First checks: Check alcohol level and supply tubing/connections. Refill as needed; persistent errors require further diagnosis. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E94Printer | Printer not printing | Check paper supply/installation and printer release lever.E94: cause & checksE94 — Printer not printingPossible cause: Paper empty or printer release mechanism not positioned correctly. First checks: Check paper supply/installation and printer release lever. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
| E95Chemicals | Detergent unavailable or not feeding | Replace/refill the approved detergent supply and restart. If detergent is present but E95 returns, inspect/prime the detergent feed system.E95: cause & checksE95 — Detergent unavailable or not feedingPossible cause: Detergent depleted or detergent supply/priming problem. First checks: Replace/refill the approved detergent supply and restart. If detergent is present but E95 returns, inspect/prime the detergent feed system. Follow the applicable Olympus operation manual for the complete procedure. If the error persists, request service support. |
No matching codes. Try a shorter search, choose All, or use Reset.
Need help with an OER-Pro error?
Contact MedServiceRepair about service support. Include the displayed error code and the checks already performed.