"A" Pre Filter A1501

srf01-a-filter-steris_1476344708

Product Code: SRF01

MFG Product ID : A1501

( Each )



Specifications

Steris

“A” Pre Filter: 2.5 micron compatible with Steris System 1E

SRF01 is a compatible filter for OEM Part A1501. This is referred to as the “A” Pre filter for Steris System 1E and Reliance Endoscope Reprocessors.

 

Instructions for changing the Pre-Filters:

1. Turn OFF water supply.

2. Exhaust air. Press red button located on top of pre-filter to exhaust air.

3. Remove housing. Unscrew housing bowl from housing body (counterclockwise).

4. Remove filter. Remove and discard old filter, rinse housing bowl with clear water.

5a. Install new filter: For A filters or single filter housing assemblies: Place the filter cartridge into the housing bowl with black O-ring facing up.

5b. Install new filter: For B filters: Install filter with black O-ring up, placing filter cartridge onto housing body.

6. Replace filter bowl. Screw filter bowl onto filter housing (clockwise).

7. Turn ON water. Restore water to filter assembly and SYSTEM 1.

8. Exhaust air. Exhaust air from both pre-filters. Press red button located on top of pre-filter housing to exhaust air.

*The above is adapted from the System 1 Maintenance Manual.

 

Steris® is a registered trademark of Steris Corporation™. MedService Repair is not affiliated with Steris® and any and all references to Steris® on this website are intended to be for informational purposes only. The products on this site have not been manufactured or approved by Steris®.

MedService Repair is your #1 source for OEM compatible AER filters at the lowest possible prices. We have years of experience with AER filters! We’ve been servicing and repairing AERs for decades which gives us firsthand knowledge of what works and what doesn’t. That knowledge becomes your advantage as we’ve spent a great deal of time considering the water filters and maintenance products we offer our clients and use ourselves repairing and maintaining AERs for medical offices throughout the United States. By creating a free account today you’ll have access to our member only special pricing on AER filters and many other parts and service items to keep your endoscope reprocessor in excellent working condition.

For years, buying the manufacturer’s OEM AER filters was the only logical solution as they were the highest quality parts. This was the was case for all the most popular AER brands; Medivators AER Filters, Cantel AER Filters, ASP AER Filters, Johnson & Johnson AER Filters, Olympus AER Filters, Custom Ultrasonics AER Filters, Steris AER Filters.

And while OEM quality has not suffered, today other options exist. Manufacturer compatible AER filters are available from reputable sources you can trust at highly discounted prices. The savings available is instrumental to running medical practices efficiently especially when the quality of the AER replacement and maintenance parts themselves meets or exceeds the manufacturer’s own high quality water filters.

The manufacturing processes of AER filters have evolved considerably over the years. Automation and technology advances in the manufacturing processes play a large role in this. For instance, the quality of the side seam welding has improved dramatically with new equipment that has helped these competing AER filter manufacturers meet or exceed the OEM manufacturer’s own strict guidelines. The glues and additives that used to be used and were a source of common issues have been replaced with ultrasonic thermal welding which completely ensured the integrity of the seams.

All this to say that if you haven’t considered aftermarket OEM compatible AER Filters as the solution for your medical practice’s consumable maintenance items you may want to consider it. If you have any questions at all about AER Filters, please call us today at the number below. We’ll be happy to answer any questions and make recommendations.