DSD-EDGE Error Codes & Troubleshooting

Medivators DSD-EDGE Error Codes & Troubleshooting Guide

Look up alarms and display messages for the Medivators DSD-EDGE automated endoscope reprocessor. The guide includes all 48 error-message appendix entries in STERIS manual 4002160/C plus 16 related prompts, log entries, legacy messages and symptoms.

Before troubleshooting: Record the exact message, station (A or B) and cycle phase. Keep an affected endoscope out of patient use until the applicable manufacturer instructions confirm successful reprocessing.

DSD-EDGE error messages and troubleshooting reference
Message or symptomWhat it meansTroubleshooting and next steps
A Dose ErrorChemistry / SSG Fault

The Part A dosing step timed out before the full delivery.

Escalate to qualified service.

A Dose Error detailsShow MoreShow Less

The Single Shot Generator (SSG) dispenses disinfectant components. This alarm identifies a timeout during delivery from the Part A dosing reservoir. An SSG A low alarm points to an earlier step, when that reservoir is being filled. The appendix specifies the point of failure but gives no component diagnosis or operator repair for this message.

STERIS 4002160/C, p. 70, p. 75, p. 76.

AbortedCycle / Prompts Status

A processing cycle was cancelled before it finished.

Keep the scope out of clinical use and follow the OEM procedure to rerun disinfection.

Aborted detailsShow MoreShow Less

The STERIS operator manual distinguishes terminating a cycle from pausing one. Termination ends processing for that scope and leaves the station in its safe mode. The manual requires another disinfection cycle for the affected endoscope before it can be released for patient use.

STERIS 4002160/C, p. 70, p. 43.

Air DisabledFlow / Air Configuration

Airflow monitoring is switched off.

Have qualified service verify the configuration.

Air Disabled detailsShow MoreShow Less

Confirm applicability to the installed unit before using the affected station.

STERIS 4002160/C, p. 70.

Attach RestrictorFlow / Air Prompt

The procedure calls for the scope simulator in place of a scope hookup.

Use the specified restrictor only in the applicable manufacturer procedure.

Attach Restrictor detailsShow MoreShow Less

STERIS 4002160/C, p. 68.

B Dose ErrorChemistry / SSG Fault

The Part B dosing step timed out before the full delivery.

Escalate to qualified service.

B Dose Error detailsShow MoreShow Less

This alarm concerns delivery from the Part B dosing reservoir. In the RAPICIDE PA workflow described in 4002160/C, Part B is the buffer. The SSG B low alarm refers to the earlier reservoir-fill step. Record the exact message so service can distinguish an interrupted dose from a low supply indication.

STERIS 4002160/C, p. 70, p. 30, p. 75, p. 76.

Bas DisabledWater / Drain Configuration

Basin-level monitoring is switched off.

Have qualified service verify the configuration.

Bas Disabled detailsShow MoreShow Less
...
Read more