Medivators DSD-201 Error Codes & Troubleshooting Guide

Find a displayed message or troubleshoot a symptom. This reference covers all 38 entries in the reviewed manual’s error-message appendix, plus 10 symptom and normal-operation references. DSD-201 messages use words and abbreviations rather than E-numbers.

48 reference entries: 38 appendix messages + 10 symptom / normal-operation entries
Message / symptomWhat it means
Troubleshooting reference & next action
AbortedCycle / Prompts Status

Cycle cancelled; processing incomplete.

Reprocess the scope before patient use.

Possible cause / context: Operator cancellation.Show MoreShow Less

An aborted processing cycle is not a completed cycle. See the interruption guidance below. During the manual’s disinfectant-loading procedure, this message can be expected; identify which operation was running.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 67, p. 18, p. 92.

Request Service Support
Air DisabledAir / Flow Configuration

Airflow monitoring disabled.

Have service verify the configuration.

Possible cause / context: Monitoring disabled.Show MoreShow Less

See the disabled-monitoring note below.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92.

Request Service Support
Bas DisabledWater / Drain Configuration

Basin-level monitoring disabled.

Have service verify the configuration.

Possible cause / context: Monitoring disabled.Show MoreShow Less

See the disabled-monitoring note below.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92.

Request Service Support
Bas Sen ErrWater / Drain Fault

Basin reads full at startup.

Check sensor droplets and kinked lines.

Possible cause / context: Wet sensor, retained liquid or kinked return/drain line.Show MoreShow Less

Remove sensor droplets and correct accessible line kinks. Escalate retained fluid or recurring errors.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 92.

Request Service Support
Basin TempTemperature Fault

Basin below programmed minimum temperature.

Refer persistent temperature errors to service.

Possible cause / context: Definition does not identify a failed component.Show MoreShow Less

Do not lower the required processing temperature to clear the message. Verify the applicable disinfectant instructions.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92, p. 8.

Request Service Support
Cancel?Cycle / Prompts Prompt

Cycle-cancellation confirmation.

Confirm only if cancellation is intended.

Possible cause / context: Cancellation requested.Show MoreShow Less

Cancelling processing requires reprocessing before patient use.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92, p. 67.

Request Service Support
Clear Sys Alarm?Electrical / System Prompt

System-alarm clearing prompt.

Leave diagnostic alarm clearing to service.

Possible cause / context: Diagnostic function selected.Show MoreShow Less

Clearing a message is not evidence that its cause has been corrected.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92.

Request Service Support
Dis DisabledDisinfectant Status

Disinfectant-count warning acknowledged.

Continue required disinfectant testing.

Possible cause / context: Warning acknowledgement.Show MoreShow Less

This is different from a disabled safety sensor. An acknowledgement does not establish disinfectant potency.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92, p. 8.

Request Service Support
Dis ExpiredDisinfectant Warning

Disinfectant cycle-count limit reached.

Replace disinfectant per the applicable instructions.

Possible cause / context: Count limit reached; possibly incorrect count setting.Show MoreShow Less

Do not raise the limit or reset the count to extend use. Ask service to resolve suspected setting errors. See the source discrepancy note below.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 92, p. 8.

Request Service Support
Dis WarningAlso: Dis WarnDisinfectant Warning

Ten cycles remain before count limit.

Check potency and plan solution replacement.

Possible cause / context: Disinfectant use count nearing limit.Show MoreShow Less

Acknowledgement does not replace the required pre-cycle MRC test.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 92, p. 8.

Request Service Support
Flash High ErrorElectrical / System Fault

High-code flash integrity check failed.

Record the message and contact service.

Possible cause / context: Component-level cause not specified.Show MoreShow Less

A memory integrity error needs technical assessment; this reference does not supply a reset sequence.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92.

Request Service Support
Flash Low ErrorElectrical / System Fault

Low-code flash integrity check failed.

Record the message and contact service.

Possible cause / context: Component-level cause not specified.Show MoreShow Less

A memory integrity error needs technical assessment; this reference does not supply a reset sequence.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92.

Request Service Support
Flo DisabledAir / Flow Configuration

Fluid-flow monitoring disabled.

Have service verify the configuration.

Possible cause / context: Monitoring disabled.Show MoreShow Less

See the disabled-monitoring note below.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92.

Request Service Support
Flow Sen ErrAir / Flow Fault

Flow reported at startup.

Use the manual’s STOP / ADD AIR check.

Possible cause / context: Flow sensor stuck on.Show MoreShow Less

See page 88 for the operator check; recurring errors need service.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 92.

Request Service Support
Hi DisabledDisinfectant Configuration

High-reservoir-level monitoring disabled.

Have service verify the configuration.

Possible cause / context: Monitoring disabled.Show MoreShow Less

See the disabled-monitoring note below.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92.

Request Service Support
High Dis ResDisinfectant Fault

Disinfectant reservoir overfull.

Correct level; check disinfectant potency.

Possible cause / context: Overfill or water entering reservoir.Show MoreShow Less

Use the prescribed draining procedure. Suspected dilution or recurrence needs service.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 89, p. 92, p. 8.

Request Service Support
Lid ajarLid / Leak Test Fault

Lid open.

Close lid; check for an obstruction.

Possible cause / context: Lid not closed.Show MoreShow Less

At startup, close it fully before resuming. Persistent indication needs service; do not bypass the lid sensor.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 92.

Request Service Support
Lid DisabledLid / Leak Test Configuration

Lid-open monitoring disabled.

Have service verify the configuration.

Possible cause / context: Monitoring disabled.Show MoreShow Less

See the disabled-monitoring note below.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92.

Request Service Support
Low AlcoholChemicals Warning

Alcohol supply low.

Replenish using the applicable procedure.

Possible cause / context: Low supply level.Show MoreShow Less

Use the specified alcohol and handling procedure; persistent indication needs service.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92, p. 68.

Request Service Support
Low ChamberWater / Drain Fault

Basin-fill time exceeded.

Identify the phase before checking supplies.

Possible cause / context: Restricted supply/filter or insufficient air; harness issue possible.Show MoreShow Less

Disinfect: check disinfectant filter and connector debris. Rinse: check water supply and filters. Ask facilities/service to assess pressure and wiring.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 89, p. 92.

Request Service Support
Low Det.Also: Low Det Low DetergentChemicals Warning

Detergent supply low.

Clean/refill using the prescribed detergent.

Possible cause / context: Low supply level.Show MoreShow Less

Use the approved detergent procedure; persistent indication needs service.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92, p. 68, p. 8.

Request Service Support
Low Dis ResDisinfectant Fault

Disinfectant reservoir underfilled.

Cancel; restore the prescribed reservoir level.

Possible cause / context: Low solution level or unit not level.Show MoreShow Less

Follow the approved loading procedure and verify potency. Suspected levelling issues need service.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 89, p. 92, p. 8.

Request Service Support
Low DisabledDisinfectant Configuration

Low-reservoir-level monitoring disabled.

Have service verify the configuration.

Possible cause / context: Monitoring disabled.Show MoreShow Less

See the disabled-monitoring note below.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93.

Request Service Support
Low Res ErrDisinfectant Fault

Reservoir-empty indication timed out during dumping.

Contact service for assessment.

Possible cause / context: Specific component cause not established by the definition.Show MoreShow Less

Record whether liquid remains and which station was dumping. Do not bypass the level sensor.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93.

Request Service Support
No Air FlowAir / Flow Fault

Airflow not detected.

Check hookup and air-filter connections.

Possible cause / context: Kink/disconnection, blocked filter, compressor or supply problem.Show MoreShow Less

Replace a blocked filter per the manual. Compressor or external-pressure problems need support.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 89, p. 93.

Request Service Support
No Fluid FloAlso: No Fluid FlowAir / Flow Fault

Fluid flow not detected.

Check the hookup and identify the cycle phase.

Possible cause / context: Disconnection, restriction, filter problem or supply fault.Show MoreShow Less

Disinfect: check filter connection/priming and hookup kinks. Flush/rinse: check water supply, filters and trapped air. A blocked scope channel needs repair; wiring checks belong to service.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 89, p. 90, p. 93.

Request Service Support
NVRAM ErrElectrical / System Fault

Nonvolatile-memory error.

Record the message and contact service.

Possible cause / context: Specific hardware cause not established.Show MoreShow Less

Diagnostic clearing is not an operator repair or proof of recovery.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93.

Request Service Support
Power onElectrical / System Status

Power interrupted during processing.

Review the cycle record and completion.

Possible cause / context: Power loss.Show MoreShow Less

The manual describes automatic cycle restart after power returns. Do not equate restored power with completed processing.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 67, p. 93.

Request Service Support
RAM ErrorElectrical / System Fault

RAM error.

Record the message and contact service.

Possible cause / context: Specific hardware cause not established.Show MoreShow Less

Diagnostic clearing is not an operator repair or proof of recovery.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93.

Request Service Support
Res T HighTemperature Fault

Reservoir above maximum temperature.

Contact service; retain the temperature reading.

Possible cause / context: Component-level cause not specified.Show MoreShow Less

Do not change processing limits to suppress this message.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93.

Request Service Support
Res T LowTemperature Fault

Reservoir below minimum temperature.

Verify warm-up; escalate a persistent error.

Possible cause / context: Reservoir insufficiently warm; component cause unconfirmed.Show MoreShow Less

Check heater status and follow the applicable startup/warm-up instructions. Do not lower the required temperature.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93, p. 18, p. 8.

Request Service Support
Reset Alarms?Cycle / Prompts Prompt

Alarm-reset confirmation.

Resolve the underlying fault first.

Possible cause / context: Reset requested.Show MoreShow Less

An acknowledgement does not prove a successful processing cycle.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93.

Request Service Support
Sheath FailLid / Leak Test Fault

Optional leak test failed.

Follow the failed-leak-test procedure; assess scope integrity.

Possible cause / context: Scope/connector leak or missing adapter.Show MoreShow Less

Page 64 distinguishes an initial failure from a small leak detected later. Do not use a suspected leaking scope on a patient. Follow both manufacturers’ instructions and obtain service guidance; see the source discrepancy below.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 64, p. 90, p. 93.

Request Service Support
Shth Sen ErrAlso: Sheath Sensor ErrorLid / Leak Test Fault

Leak-test pressure present at startup.

Check the leak-test setup using page 90.

Possible cause / context: Retained pressure or incorrect option configuration.Show MoreShow Less

Follow the specified hookup depressurization/reconnection procedure. An enabled but absent tester requires service configuration.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 90, p. 93.

Request Service Support
SNVRAM ErrElectrical / System Fault

Short nonvolatile-memory error.

Record the message and contact service.

Possible cause / context: Specific hardware cause not established.Show MoreShow Less

Diagnostic clearing is not an operator repair or proof of recovery.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93.

Request Service Support
Sta Not IdleCycle / Prompts Prompt

Operation requires an idle station.

Wait for the station to become idle.

Possible cause / context: Another operation active.Show MoreShow Less

Check that the intended station is selected.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93.

Request Service Support
Sth DisabledLid / Leak Test Configuration

Optional leak tester disabled.

Verify the installed option and facility procedure.

Possible cause / context: Tester disabled.Show MoreShow Less

The automated test does not replace the scope manufacturer’s manual leak test. Ask service about an unexpected setting.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93, p. 64.

Request Service Support
Time ErrElectrical / System Fault

Timebase error.

Record the message and contact service.

Possible cause / context: Specific hardware cause not established.Show MoreShow Less

Diagnostic clearing is not an operator repair or proof of recovery.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93.

Request Service Support
Unit will not startElectrical / System Symptom

No operating response.

Check supply power; escalate recurring trips.

Possible cause / context: Disconnected power or tripped protection.Show MoreShow Less

Follow the manual’s GFCI/breaker checks. If protection cannot reset or trips again, obtain service.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 86.

Request Service Support
Slow water fillWater / Drain Symptom

Basin fills unusually slowly.

Check supply and filter condition.

Possible cause / context: Restricted water/air, filter blockage or pressure/harness issue.Show MoreShow Less

Facilities/service should assess pressure and wiring. Do not adjust factory regulators as a workaround.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 86, p. 7.

Request Service Support
Disinfectant will not returnDisinfectant Symptom

Solution remains above the reservoir.

Check basin screen and return-line routing.

Possible cause / context: Blocked screen or kinked return line.Show MoreShow Less

Use the prescribed cleaning procedure; recurring drainage failure needs service.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 86.

Request Service Support
Continuous water from hookupsWater / Drain Symptom

Water keeps flowing.

Shut off water and contact service.

Possible cause / context: Water valve fails to seal.Show MoreShow Less

Do not continue operating with uncontrolled flow.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87.

Request Service Support
Basin will not drainWater / Drain Symptom

Flush/rinse water remains in basin.

Check screen and external drain routing.

Possible cause / context: Drain too high, kinked/blocked tubing or blocked screen.Show MoreShow Less

Drain must be below the outlet. Have qualified personnel correct routing or blockage.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87.

Request Service Support
No air after cycleAir / Flow Symptom

Expected final air phase absent.

Have the selected program checked.

Possible cause / context: Incorrect air-program setting.Show MoreShow Less

See Setup 5; maintain the validated processing parameters.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87.

Request Service Support
LCD unreadableElectrical / System Symptom

Display cannot be read.

Contact service.

Possible cause / context: Contrast setting or display failure.Show MoreShow Less

Do not infer completion from an unreadable display.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87.

Request Service Support
Printing unreadablePrinter Symptom

Printout faint or illegible.

Check ribbon on ribbon-equipped printers.

Possible cause / context: Ribbon unused for a period or worn.Show MoreShow Less

Advance or replace ribbon per the manual. Thermal-printer versions require their own instructions.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87.

Request Service Support
Printer will not operatePrinter Symptom

No printed output.

Check printer power, paper and connections.

Possible cause / context: Power/switch, connection, paper/ribbon jam or printer failure.Show MoreShow Less

Follow printer-specific maintenance; escalate hardware faults.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87.

Request Service Support
Brief delay before disinfectionLid / Leak Test Normal

Optional initial leak test in progress.

Allow the initial test to finish.

Possible cause / context: Normal 40-second test sequence.Show MoreShow Less

A delay with no fault during this test is expected. A failed test requires the Sheath Fail response.

MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 86, p. 64.

Request Service Support

Interrupted or aborted cycle?

The manual distinguishes a pause from an abort: a corrected interruption may resume, while an aborted processing cycle requires reprocessing. It describes automatic restart after power restoration. Confirm completed processing before releasing the scope. See p. 67.

Important distinctions

Disabled monitoring: these entries describe disabled functions, not a confirmed failed sensor. As a precaution, have unexpected configuration notices reviewed by qualified service before clinical use. Never disable monitoring to clear a fault. Optional leak-test availability must be checked for the installed unit.

Disinfectant: test the minimum required concentration (MRC) before each reprocessing cycle using the specified strips. Replace solution below MRC and comply with the disinfectant’s use limits. Temperature and contact time depend on the applicable disinfectant instructions. See p. 8.

Pressure: this guide deliberately does not instruct operators to adjust regulators. The manual identifies factory presets and directs users to technical support. See p. 7.

Source differences that affect troubleshooting

Dis Expired: page 88 describes checking potency and replacement “if necessary”; page 92 says the maximum count prevents further cycles before solution replacement. This guide uses the more conservative replacement direction and refers count-setting questions to service.

Sheath Fail: page 64 distinguishes initial failure (abort) from a small leak detected during processing. Page 90’s troubleshooting table gives broader continuation wording. This guide does not turn that into a blanket instruction to continue processing a leaking scope. Obtain model-specific guidance and follow the endoscope manufacturer’s failed-leak-test instructions.

Need help with a DSD-201 message?

Include the exact message, station, cycle phase and checks already performed.

Request Service Support

Source edition reviewed: MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L, © 2016; error-message appendix pp. 92–93, troubleshooting pp. 86–90, and related operator sections. Message definitions were cross-checked against the manufacturer’s service-manual appendix (publicly hosted copy, pp. 148–149). Reviewed October 22, 2026.

Scope: manufacturer-authored manuals accessed through third-party hosts. The former Medivators PDF link redirected to STERIS during research. Current OEM revision and applicability to every software/configuration have not been confirmed. This is a concise reference, not the full operating or service procedure, and does not cover DSD EDGE. Causes are listed only where supported; otherwise the guide refers the issue to service.