Medivators DSD-201 Error Codes & Troubleshooting Guide
Find a displayed message or troubleshoot a symptom. This reference covers all 38 entries in the reviewed manual’s error-message appendix, plus 10 symptom and normal-operation references. DSD-201 messages use words and abbreviations rather than E-numbers.
Before acting: record Station A/B and the cycle phase. Use the applicable manufacturer instructions and trained personnel. Acknowledging an alarm does not confirm successful reprocessing.
| Message / symptom | What it means |
Troubleshooting reference & next action
|
|---|---|---|
| AbortedCycle / Prompts | Status
Cycle cancelled; processing incomplete. |
Reprocess the scope before patient use. Possible cause / context: Operator cancellation.Show MoreShow LessAn aborted processing cycle is not a completed cycle. See the interruption guidance below. During the manual’s disinfectant-loading procedure, this message can be expected; identify which operation was running. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 67, p. 18, p. 92. Request Service Support |
| Air DisabledAir / Flow | Configuration
Airflow monitoring disabled. |
Have service verify the configuration. Possible cause / context: Monitoring disabled.Show MoreShow LessSee the disabled-monitoring note below. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92. Request Service Support |
| Bas DisabledWater / Drain | Configuration
Basin-level monitoring disabled. |
Have service verify the configuration. Possible cause / context: Monitoring disabled.Show MoreShow LessSee the disabled-monitoring note below. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92. Request Service Support |
| Bas Sen ErrWater / Drain | Fault
Basin reads full at startup. |
Check sensor droplets and kinked lines. Possible cause / context: Wet sensor, retained liquid or kinked return/drain line.Show MoreShow LessRemove sensor droplets and correct accessible line kinks. Escalate retained fluid or recurring errors. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 92. Request Service Support |
| Basin TempTemperature | Fault
Basin below programmed minimum temperature. |
Refer persistent temperature errors to service. Possible cause / context: Definition does not identify a failed component.Show MoreShow LessDo not lower the required processing temperature to clear the message. Verify the applicable disinfectant instructions. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92, p. 8. Request Service Support |
| Cancel?Cycle / Prompts | Prompt
Cycle-cancellation confirmation. |
Confirm only if cancellation is intended. Possible cause / context: Cancellation requested.Show MoreShow LessCancelling processing requires reprocessing before patient use. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92, p. 67. Request Service Support |
| Clear Sys Alarm?Electrical / System | Prompt
System-alarm clearing prompt. |
Leave diagnostic alarm clearing to service. Possible cause / context: Diagnostic function selected.Show MoreShow LessClearing a message is not evidence that its cause has been corrected. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92. Request Service Support |
| Dis DisabledDisinfectant | Status
Disinfectant-count warning acknowledged. |
Continue required disinfectant testing. Possible cause / context: Warning acknowledgement.Show MoreShow LessThis is different from a disabled safety sensor. An acknowledgement does not establish disinfectant potency. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92, p. 8. Request Service Support |
| Dis ExpiredDisinfectant | Warning
Disinfectant cycle-count limit reached. |
Replace disinfectant per the applicable instructions. Possible cause / context: Count limit reached; possibly incorrect count setting.Show MoreShow LessDo not raise the limit or reset the count to extend use. Ask service to resolve suspected setting errors. See the source discrepancy note below. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 92, p. 8. Request Service Support |
| Dis WarningAlso: Dis WarnDisinfectant | Warning
Ten cycles remain before count limit. |
Check potency and plan solution replacement. Possible cause / context: Disinfectant use count nearing limit.Show MoreShow LessAcknowledgement does not replace the required pre-cycle MRC test. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 92, p. 8. Request Service Support |
| Flash High ErrorElectrical / System | Fault
High-code flash integrity check failed. |
Record the message and contact service. Possible cause / context: Component-level cause not specified.Show MoreShow LessA memory integrity error needs technical assessment; this reference does not supply a reset sequence. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92. Request Service Support |
| Flash Low ErrorElectrical / System | Fault
Low-code flash integrity check failed. |
Record the message and contact service. Possible cause / context: Component-level cause not specified.Show MoreShow LessA memory integrity error needs technical assessment; this reference does not supply a reset sequence. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92. Request Service Support |
| Flo DisabledAir / Flow | Configuration
Fluid-flow monitoring disabled. |
Have service verify the configuration. Possible cause / context: Monitoring disabled.Show MoreShow LessSee the disabled-monitoring note below. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92. Request Service Support |
| Flow Sen ErrAir / Flow | Fault
Flow reported at startup. |
Use the manual’s STOP / ADD AIR check. Possible cause / context: Flow sensor stuck on.Show MoreShow LessSee page 88 for the operator check; recurring errors need service. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 92. Request Service Support |
| Hi DisabledDisinfectant | Configuration
High-reservoir-level monitoring disabled. |
Have service verify the configuration. Possible cause / context: Monitoring disabled.Show MoreShow LessSee the disabled-monitoring note below. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92. Request Service Support |
| High Dis ResDisinfectant | Fault
Disinfectant reservoir overfull. |
Correct level; check disinfectant potency. Possible cause / context: Overfill or water entering reservoir.Show MoreShow LessUse the prescribed draining procedure. Suspected dilution or recurrence needs service. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 89, p. 92, p. 8. Request Service Support |
| Lid ajarLid / Leak Test | Fault
Lid open. |
Close lid; check for an obstruction. Possible cause / context: Lid not closed.Show MoreShow LessAt startup, close it fully before resuming. Persistent indication needs service; do not bypass the lid sensor. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 92. Request Service Support |
| Lid DisabledLid / Leak Test | Configuration
Lid-open monitoring disabled. |
Have service verify the configuration. Possible cause / context: Monitoring disabled.Show MoreShow LessSee the disabled-monitoring note below. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92. Request Service Support |
| Low AlcoholChemicals | Warning
Alcohol supply low. |
Replenish using the applicable procedure. Possible cause / context: Low supply level.Show MoreShow LessUse the specified alcohol and handling procedure; persistent indication needs service. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92, p. 68. Request Service Support |
| Low ChamberWater / Drain | Fault
Basin-fill time exceeded. |
Identify the phase before checking supplies. Possible cause / context: Restricted supply/filter or insufficient air; harness issue possible.Show MoreShow LessDisinfect: check disinfectant filter and connector debris. Rinse: check water supply and filters. Ask facilities/service to assess pressure and wiring. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 88, p. 89, p. 92. Request Service Support |
| Low Det.Also: Low Det Low DetergentChemicals | Warning
Detergent supply low. |
Clean/refill using the prescribed detergent. Possible cause / context: Low supply level.Show MoreShow LessUse the approved detergent procedure; persistent indication needs service. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 92, p. 68, p. 8. Request Service Support |
| Low Dis ResDisinfectant | Fault
Disinfectant reservoir underfilled. |
Cancel; restore the prescribed reservoir level. Possible cause / context: Low solution level or unit not level.Show MoreShow LessFollow the approved loading procedure and verify potency. Suspected levelling issues need service. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 89, p. 92, p. 8. Request Service Support |
| Low DisabledDisinfectant | Configuration
Low-reservoir-level monitoring disabled. |
Have service verify the configuration. Possible cause / context: Monitoring disabled.Show MoreShow LessSee the disabled-monitoring note below. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93. Request Service Support |
| Low Res ErrDisinfectant | Fault
Reservoir-empty indication timed out during dumping. |
Contact service for assessment. Possible cause / context: Specific component cause not established by the definition.Show MoreShow LessRecord whether liquid remains and which station was dumping. Do not bypass the level sensor. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93. Request Service Support |
| No Air FlowAir / Flow | Fault
Airflow not detected. |
Check hookup and air-filter connections. Possible cause / context: Kink/disconnection, blocked filter, compressor or supply problem.Show MoreShow LessReplace a blocked filter per the manual. Compressor or external-pressure problems need support. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 89, p. 93. Request Service Support |
| No Fluid FloAlso: No Fluid FlowAir / Flow | Fault
Fluid flow not detected. |
Check the hookup and identify the cycle phase. Possible cause / context: Disconnection, restriction, filter problem or supply fault.Show MoreShow LessDisinfect: check filter connection/priming and hookup kinks. Flush/rinse: check water supply, filters and trapped air. A blocked scope channel needs repair; wiring checks belong to service. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 89, p. 90, p. 93. Request Service Support |
| NVRAM ErrElectrical / System | Fault
Nonvolatile-memory error. |
Record the message and contact service. Possible cause / context: Specific hardware cause not established.Show MoreShow LessDiagnostic clearing is not an operator repair or proof of recovery. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93. Request Service Support |
| Power onElectrical / System | Status
Power interrupted during processing. |
Review the cycle record and completion. Possible cause / context: Power loss.Show MoreShow LessThe manual describes automatic cycle restart after power returns. Do not equate restored power with completed processing. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 67, p. 93. Request Service Support |
| RAM ErrorElectrical / System | Fault
RAM error. |
Record the message and contact service. Possible cause / context: Specific hardware cause not established.Show MoreShow LessDiagnostic clearing is not an operator repair or proof of recovery. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93. Request Service Support |
| Res T HighTemperature | Fault
Reservoir above maximum temperature. |
Contact service; retain the temperature reading. Possible cause / context: Component-level cause not specified.Show MoreShow LessDo not change processing limits to suppress this message. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93. Request Service Support |
| Res T LowTemperature | Fault
Reservoir below minimum temperature. |
Verify warm-up; escalate a persistent error. Possible cause / context: Reservoir insufficiently warm; component cause unconfirmed.Show MoreShow LessCheck heater status and follow the applicable startup/warm-up instructions. Do not lower the required temperature. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93, p. 18, p. 8. Request Service Support |
| Reset Alarms?Cycle / Prompts | Prompt
Alarm-reset confirmation. |
Resolve the underlying fault first. Possible cause / context: Reset requested.Show MoreShow LessAn acknowledgement does not prove a successful processing cycle. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93. Request Service Support |
| Sheath FailLid / Leak Test | Fault
Optional leak test failed. |
Follow the failed-leak-test procedure; assess scope integrity. Possible cause / context: Scope/connector leak or missing adapter.Show MoreShow LessPage 64 distinguishes an initial failure from a small leak detected later. Do not use a suspected leaking scope on a patient. Follow both manufacturers’ instructions and obtain service guidance; see the source discrepancy below. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 64, p. 90, p. 93. Request Service Support |
| Shth Sen ErrAlso: Sheath Sensor ErrorLid / Leak Test | Fault
Leak-test pressure present at startup. |
Check the leak-test setup using page 90. Possible cause / context: Retained pressure or incorrect option configuration.Show MoreShow LessFollow the specified hookup depressurization/reconnection procedure. An enabled but absent tester requires service configuration. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 90, p. 93. Request Service Support |
| SNVRAM ErrElectrical / System | Fault
Short nonvolatile-memory error. |
Record the message and contact service. Possible cause / context: Specific hardware cause not established.Show MoreShow LessDiagnostic clearing is not an operator repair or proof of recovery. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93. Request Service Support |
| Sta Not IdleCycle / Prompts | Prompt
Operation requires an idle station. |
Wait for the station to become idle. Possible cause / context: Another operation active.Show MoreShow LessCheck that the intended station is selected. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93. Request Service Support |
| Sth DisabledLid / Leak Test | Configuration
Optional leak tester disabled. |
Verify the installed option and facility procedure. Possible cause / context: Tester disabled.Show MoreShow LessThe automated test does not replace the scope manufacturer’s manual leak test. Ask service about an unexpected setting. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93, p. 64. Request Service Support |
| Time ErrElectrical / System | Fault
Timebase error. |
Record the message and contact service. Possible cause / context: Specific hardware cause not established.Show MoreShow LessDiagnostic clearing is not an operator repair or proof of recovery. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 93. Request Service Support |
| Unit will not startElectrical / System | Symptom
No operating response. |
Check supply power; escalate recurring trips. Possible cause / context: Disconnected power or tripped protection.Show MoreShow LessFollow the manual’s GFCI/breaker checks. If protection cannot reset or trips again, obtain service. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 86. Request Service Support |
| Slow water fillWater / Drain | Symptom
Basin fills unusually slowly. |
Check supply and filter condition. Possible cause / context: Restricted water/air, filter blockage or pressure/harness issue.Show MoreShow LessFacilities/service should assess pressure and wiring. Do not adjust factory regulators as a workaround. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 86, p. 7. Request Service Support |
| Disinfectant will not returnDisinfectant | Symptom
Solution remains above the reservoir. |
Check basin screen and return-line routing. Possible cause / context: Blocked screen or kinked return line.Show MoreShow LessUse the prescribed cleaning procedure; recurring drainage failure needs service. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 86. Request Service Support |
| Continuous water from hookupsWater / Drain | Symptom
Water keeps flowing. |
Shut off water and contact service. Possible cause / context: Water valve fails to seal.Show MoreShow LessDo not continue operating with uncontrolled flow. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87. Request Service Support |
| Basin will not drainWater / Drain | Symptom
Flush/rinse water remains in basin. |
Check screen and external drain routing. Possible cause / context: Drain too high, kinked/blocked tubing or blocked screen.Show MoreShow LessDrain must be below the outlet. Have qualified personnel correct routing or blockage. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87. Request Service Support |
| No air after cycleAir / Flow | Symptom
Expected final air phase absent. |
Have the selected program checked. Possible cause / context: Incorrect air-program setting.Show MoreShow LessSee Setup 5; maintain the validated processing parameters. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87. Request Service Support |
| LCD unreadableElectrical / System | Symptom
Display cannot be read. |
Contact service. Possible cause / context: Contrast setting or display failure.Show MoreShow LessDo not infer completion from an unreadable display. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87. Request Service Support |
| Printing unreadablePrinter | Symptom
Printout faint or illegible. |
Check ribbon on ribbon-equipped printers. Possible cause / context: Ribbon unused for a period or worn.Show MoreShow LessAdvance or replace ribbon per the manual. Thermal-printer versions require their own instructions. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87. Request Service Support |
| Printer will not operatePrinter | Symptom
No printed output. |
Check printer power, paper and connections. Possible cause / context: Power/switch, connection, paper/ribbon jam or printer failure.Show MoreShow LessFollow printer-specific maintenance; escalate hardware faults. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 87. Request Service Support |
| Brief delay before disinfectionLid / Leak Test | Normal
Optional initial leak test in progress. |
Allow the initial test to finish. Possible cause / context: Normal 40-second test sequence.Show MoreShow LessA delay with no fault during this test is expected. A failed test requires the Sheath Fail response. MEDIVATORS DSD-201 User Manual, MM03-0046 Rev L: p. 86, p. 64. Request Service Support |
No matching entries. Try a shorter search, choose All, or use Reset.
Interrupted or aborted cycle?
The manual distinguishes a pause from an abort: a corrected interruption may resume, while an aborted processing cycle requires reprocessing. It describes automatic restart after power restoration. Confirm completed processing before releasing the scope. See p. 67.
Important distinctions
Disabled monitoring: these entries describe disabled functions, not a confirmed failed sensor. As a precaution, have unexpected configuration notices reviewed by qualified service before clinical use. Never disable monitoring to clear a fault. Optional leak-test availability must be checked for the installed unit.
Disinfectant: test the minimum required concentration (MRC) before each reprocessing cycle using the specified strips. Replace solution below MRC and comply with the disinfectant’s use limits. Temperature and contact time depend on the applicable disinfectant instructions. See p. 8.
Pressure: this guide deliberately does not instruct operators to adjust regulators. The manual identifies factory presets and directs users to technical support. See p. 7.
Source differences that affect troubleshooting
Dis Expired: page 88 describes checking potency and replacement “if necessary”; page 92 says the maximum count prevents further cycles before solution replacement. This guide uses the more conservative replacement direction and refers count-setting questions to service.
Sheath Fail: page 64 distinguishes initial failure (abort) from a small leak detected during processing. Page 90’s troubleshooting table gives broader continuation wording. This guide does not turn that into a blanket instruction to continue processing a leaking scope. Obtain model-specific guidance and follow the endoscope manufacturer’s failed-leak-test instructions.
Need help with a DSD-201 message?
Include the exact message, station, cycle phase and checks already performed.