Olympus OER-Pro Error Codes & Troubleshooting Guide

Directions

Step 1: Confirm applicable instructions for your unit with Olympus.Professional registration and sign-in may be required.

Step 2: Find an error code below, review its possible cause, and see what to check first. This reference covers 39 OER-Pro codes, with expandable guidance for routine troubleshooting checks.

Search Olympus OER-Pro error codes

OER-Pro error-code chart · 39 codes
CodeWhat it meansTroubleshooting reference & next action
E00Electrical / System Cycle/process interrupted Wait for automatic error processing to finish. See the manual for optional air-purge or rinse procedures.

Possible Cause: STOP/RESET pressed during operation.

E01Water / Drain Basin takes too long to fill with water Check the faucet, supply hose and filters.
Possible Cause: Partly closed faucet, leaking or obstructed water supply.Show MoreShow Less — E01 troubleshooting reference

Troubleshooting Reference: Open the faucet fully; check for hose leakage. The manual specifies water-filter replacement and inlet-mesh cleaning.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 277.

Request Service Support
E02Water / Drain Cleaning fluid will not drain Check the drain hose and basin drain filter.
Possible Cause: Drain-hose problem or obstructed basin drain mesh.Show MoreShow Less — E02 troubleshooting reference

Troubleshooting Reference: Check hose installation, clean the drain mesh, then follow the manual’s closed-lid air-purge procedure.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 277.

Request Service Support
E04Water / Drain Cleaning-fluid level drops unexpectedly Contact Olympus.

Possible Cause: Internal equipment fault.

E05Water / Drain Cleaning-fluid level too high Contact Olympus; close the faucet if water flow will not stop.

Possible Cause: Fluid-level sensor irregularity.

E06Sensors Middle fluid-level sensor error Clean the fluid-level sensor as instructed.
Possible Cause: Residue affecting the middle level sensor.Show MoreShow Less — E06 troubleshooting reference

Troubleshooting Reference: Use the manual’s fluid-level sensor cleaning procedure.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 278.

Request Service Support
E07Sensors Lower fluid-level sensor error Clean the fluid-level sensor as instructed.
Possible Cause: Residue affecting the lower level sensor.Show MoreShow Less — E07 troubleshooting reference

Troubleshooting Reference: Use the manual’s fluid-level sensor cleaning procedure.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 278.

Request Service Support
E11Water / Drain Too much disinfectant in disinfectant tank Check drainage and disinfectant concentration.
Possible Cause: Drain or disinfectant-return piping irregularity.Show MoreShow Less — E11 troubleshooting reference

Troubleshooting Reference: Close an open lid to allow automatic processing; check hose installation and clean the drain mesh. Test concentration and replace ineffective solution.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 278.

Request Service Support
E12Disinfectant Too little disinfectant in tank Clean circulation mesh filters; replace the disinfectant.
Possible Cause: Insufficient disinfectant in the tank.Show MoreShow Less — E12 troubleshooting reference

Troubleshooting Reference: Drain the tank completely before replacement. Do not top up used disinfectant with fresh solution.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 278.

Request Service Support
E13Water / Drain Basin fills with disinfectant unusually quickly Check drainage; follow the E13 solution-removal procedure.
Possible Cause: Incomplete drainage left liquid in the basin.Show MoreShow Less — E13 troubleshooting reference

Troubleshooting Reference: Use the complete basin/tank discharge procedure on printed page 278 before preparing new solution and restarting.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 278.

Request Service Support
E14Water / Drain Disinfectant cannot be returned to tank Check ventilation and drain/circulation mesh filters.
Possible Cause: Obstructed drain mesh or gas-filter ventilation.Show MoreShow Less — E14 troubleshooting reference

Troubleshooting Reference: Follow the collection, rinsing and solution-quantity branches on printed page 279; do not top up used disinfectant.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 279.

Request Service Support
E15Disinfectant Fluid-level sensor error Clean the level sensor; follow the manual’s solution-quantity and rinsing instructions.

Possible Cause: Middle-sensor residue or internal fault.

E16Disinfectant Basin takes too long to fill with disinfectant Contact Olympus.

Possible Cause: Internal equipment fault.

E17Disinfectant Disinfectant will not heat correctly Contact Olympus.

Possible Cause: Internal equipment fault.

E18Disinfectant Temperature sensor error Contact Olympus.

Possible Cause: Internal equipment fault.

E21Flow / Pressure Low air-purge pressure / air not flowing through scope channels Check air-filter installation.
Possible Cause: Air filter incorrectly fitted or clogged.Show MoreShow Less — E21 troubleshooting reference

Troubleshooting Reference: If correctly installed, the filter may be clogged; replace it as instructed.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 279.

Request Service Support
E22Flow / Pressure Insufficient fluid pressure Clean the circulation-port mesh filters.
Possible Cause: Obstructed circulation-port mesh filters.Show MoreShow Less — E22 troubleshooting reference

Troubleshooting Reference: Use the manual’s mesh-filter cleaning procedure.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 279.

Request Service Support
E23Flow / Pressure Excessive fluid pressure Contact Olympus.

Possible Cause: Internal equipment fault.

E31Lid / Drawer Lid open At startup, close the lid and restart. During a process, contact Olympus.
Possible Cause: Lid open at start; internal fault if detected during operation.Show MoreShow Less — E31 troubleshooting reference

Troubleshooting Reference: The manual distinguishes a startup lid error from one occurring during operation.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 280.

Request Service Support
E41Electrical / System Power lost and restored during cycle Check the connection; unplug before inspecting the cord for damage. Replace a damaged cord.

Possible Cause: Power interruption or power-cord contact failure.

E51Water / Drain Water detected leaking inside machine Close the water faucet and contact Olympus. Do not use a malfunctioning unit.

Possible Cause: Internal equipment fault.

E61Electrical / System Ultrasonic cleaning not operating Contact Olympus.

Possible Cause: Ultrasonic oscillator irregularity.

E71Disinfectant Disinfectant-bottle drawer sensor error Contact Olympus.

Possible Cause: Internal equipment fault.

E72Water / Drain Disinfectant tank will not drain If solution remains in the basin, follow E13; otherwise contact Olympus.

Possible Cause: Internal equipment fault.

E73Water / Drain Water-supply problem while preparing disinfectant Check the water supply and filters.
Possible Cause: Partly closed faucet, leaking or obstructed water supply.Show MoreShow Less — E73 troubleshooting reference

Troubleshooting Reference: Follow the water-supply checks on printed page 280; prepare disinfectant again and test its effectiveness after loading.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 280.

Request Service Support
E74Disinfectant Disinfectant from cassette bottles isn't reaching the unit Contact Olympus.

Possible Cause: Internal fault or cassette solution not reaching the tank.

E75Water / Drain Water-supply problem while preparing disinfectant Check the faucet, hose and filters.
Possible Cause: Restricted/leaking supply or temporarily reduced water delivery.Show MoreShow Less — E75 troubleshooting reference

Troubleshooting Reference: Follow printed page 281: check supply, clean inlet mesh, replace water filter, close an open lid, prepare solution and test effectiveness.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 281.

Request Service Support
E76Disinfectant Disinfectant-tank fluid-level sensor error Contact Olympus.

Possible Cause: Internal equipment fault.

E77Disinfectant Disinfectant bottle drawer not fully positioned Push the drawer fully closed.
Possible Cause: Disinfectant drawer not fully closed.Show MoreShow Less — E77 troubleshooting reference

Troubleshooting Reference: Confirm complete drawer closure as described in the manual.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 281.

Request Service Support
E78Disinfectant Cassette bottles empty Check cassette installation and whether bottles are empty.
Possible Cause: Cassette detection-switch irregularity or empty bottles installed.Show MoreShow Less — E78 troubleshooting reference

Troubleshooting Reference: Correct installation; replace empty bottles with new ones.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 281.

Request Service Support
E81Electrical / System Cycle cannot be properly controlled Contact Olympus.

Possible Cause: Internal equipment fault.

E82Electrical / System Internal machine irregularity Contact Olympus.

Possible Cause: Internal electrical-circuit irregularity.

E83Electrical / System Cycle cannot be properly controlled Contact Olympus.

Possible Cause: Internal electrical-circuit irregularity.

E84Electrical / System RFID reader malfunction Contact Olympus.

Possible Cause: Internal electrical-circuit irregularity.

E91RFID / ID ID read error Reset the reading and scan again.
Possible Cause: No ID read, or multiple IDs read.Show MoreShow Less — E91 troubleshooting reference

Troubleshooting Reference: Press STOP/RESET; confirm the ID detection indicators are unlit before rescanning.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 282.

Request Service Support
E92Flow / Pressure Leak test exceeds allowed time Repeat the leak test according to the manual.

Possible Cause: Leak testing exceeded 10 minutes.

E93Chemicals Alcohol unavailable or not feeding Use the E93 procedure for the operation in progress.
Possible Cause: Alcohol depleted/not supplied, sensor fault, or pump/piping obstruction.Show MoreShow Less — E93 troubleshooting reference

Troubleshooting Reference: During reprocessing, see printed page 117. For a separate alcohol flush, check tank quantity and follow the applicable procedure, including page 120 for recurrence.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 282.

Request Service Support
E94Printer Printer not printing Check paper installation and release-lever position.
Possible Cause: Paper exhausted or printer release lever raised.Show MoreShow Less — E94 troubleshooting reference

Troubleshooting Reference: Follow the printer-paper installation instructions and set the lever correctly.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 282.

Request Service Support
E95Chemicals Detergent unavailable or not feeding Follow the E95 procedure on printed page 125.
Possible Cause: Detergent depleted/not supplied, sensor fault, or pump/piping obstruction.Show MoreShow Less — E95 troubleshooting reference

Troubleshooting Reference: Inspect and replace the detergent tank as directed, then restart reprocessing from the beginning. If detergent remains or E95 recurs, use the manual’s specified detergent-line filling procedure; this summary is not that procedure.

Olympus OER-Pro Operation Manual, GT8357 12, §8.1, printed p. 283.

Request Service Support

Need help with an OER-Pro error?

Contact MedServiceRepair about service support. Include the displayed error code and the checks already performed.

Request Service Support

Source edition reviewed: Olympus OER-Pro Operation Manual, identifier GT8357 12, printed date code 20120111; section 8.1, printed pages 277–283. Reviewed against a publicly available third-party-hosted copy on September 28, 2026. This is not confirmation of the current OEM revision or applicability to every unit.

About this reference. Code descriptions and checks are paraphrased summaries, not the complete Olympus operating instructions. Consult the applicable OER-Pro operation manual for your unit before performing maintenance or restarting reprocessing.