Medivators DSD-EDGE Error Codes & Troubleshooting Guide
Look up alarms and display messages for the Medivators DSD-EDGE automated endoscope reprocessor. The guide includes all 48 error-message appendix entries in STERIS manual 4002160/C plus 16 related prompts, log entries, legacy messages and symptoms.
Before troubleshooting: Record the exact message, station (A or B) and cycle phase. Keep an affected endoscope out of patient use until the applicable manufacturer instructions confirm successful reprocessing.
64 of 64 entries shown
No matching entries. Try the exact display text or clear the filters.
| Message or symptom | What it means | Troubleshooting and next steps |
|---|---|---|
| A Dose ErrorChemistry / SSG | Fault
The Part A dosing step timed out before the full delivery. |
Escalate to qualified service. A Dose Error detailsShow MoreShow LessThe Single Shot Generator (SSG) dispenses disinfectant components. This alarm identifies a timeout during delivery from the Part A dosing reservoir. An SSG A low alarm points to an earlier step, when that reservoir is being filled. The appendix specifies the point of failure but gives no component diagnosis or operator repair for this message. |
| AbortedCycle / Prompts | Status
A processing cycle was cancelled before it finished. |
Keep the scope out of clinical use and follow the OEM procedure to rerun disinfection. Aborted detailsShow MoreShow LessThe STERIS operator manual distinguishes terminating a cycle from pausing one. Termination ends processing for that scope and leaves the station in its safe mode. The manual requires another disinfection cycle for the affected endoscope before it can be released for patient use. |
| Air DisabledFlow / Air | Configuration
Airflow monitoring is switched off. |
Have qualified service verify the configuration. Air Disabled detailsShow MoreShow LessConfirm applicability to the installed unit before using the affected station. STERIS 4002160/C, p. 70. |
| Attach RestrictorFlow / Air | Prompt
The procedure calls for the scope simulator in place of a scope hookup. |
Use the specified restrictor only in the applicable manufacturer procedure. Attach Restrictor detailsShow MoreShow LessSTERIS 4002160/C, p. 68. |
| B Dose ErrorChemistry / SSG | Fault
The Part B dosing step timed out before the full delivery. |
Escalate to qualified service. B Dose Error detailsShow MoreShow LessThis alarm concerns delivery from the Part B dosing reservoir. In the RAPICIDE PA workflow described in 4002160/C, Part B is the buffer. The SSG B low alarm refers to the earlier reservoir-fill step. Record the exact message so service can distinguish an interrupted dose from a low supply indication. |
| Bas DisabledWater / Drain | Configuration
Basin-level monitoring is switched off. |
Have qualified service verify the configuration. Bas Disabled detailsShow MoreShow LessSTERIS 4002160/C, p. 70. |
| Bas Sen ErrAlso search: Basin Sensor ErrorWater / Drain | Fault
The level sensor reads full at cycle start, when the basin should be empty. |
Check for retained liquid, droplets on the sensor, or a visible drain-line kink when access is safe. Bas Sen Err detailsShow MoreShow LessThe earlier Medivators operator manual lists three things to investigate at startup: droplets on the basin sensor, liquid left behind after a drain, and a kinked drain line. A visibly empty basin and a full indication provide different information from a basin that still contains liquid. Tell service which condition you observed. STERIS 4002160/C, p. 70. Earlier Medivators documentation: Medivators 50097-027 Rev E, pp. 76–80. |
| Basin does not drainWater / Drain | Symptom
Fluid remains in the basin when it should have drained. |
When safe, look for a visibly kinked drain line and contact service if fluid remains. Basin does not drain detailsShow MoreShow LessThe earlier Medivators operator manual lists a kinked external drain line or a drain positioned too high as possible contributors during flush or rinse. The STERIS specifications describe the gravity-drain arrangement. Have facilities or service verify the installation if draining remains incomplete. STERIS 4002160/C, p. 70, p. 77. Earlier Medivators documentation: Medivators 50097-027 Rev E, pp. 76–80. |
| Basin does not fillWater / Drain | Symptom
The basin is visibly underfilled or fails to fill. |
Record the cycle phase and see Low Chamber. Basin does not fill detailsShow MoreShow LessA water-supply or filter issue, an air-supply issue or a disinfectant-path restriction may apply. STERIS 4002160/C, p. 28, p. 32, p. 33, p. 70. Earlier Medivators documentation: Medivators 50097-027 Rev E, pp. 76–80, Medivators 50097-037 Rev E, pp. 126–128. |
| Basin drained?Water / Drain | Prompt
The equipment asks the operator to verify that the basin has drained. |
Confirm the basin condition before proceeding under the applicable OEM procedure. Basin drained? detailsShow MoreShow LessSTERIS 4002160/C, p. 68. |
| Basin In UseAlso search: Sta Not IdleCycle / Prompts | Status
The selected station must be idle for the requested operation. |
Verify the selected station and allow its current operation to finish as directed by the manual. Basin In Use detailsShow MoreShow LessSTERIS 4002160/C, p. 72, p. 21. Earlier Medivators documentation: Medivators 50097-027 Rev E, Medivators 50097-037 Rev E, pp. 129–130. |
| Basin TempTemperature | Fault
The basin failed to reach the programmed minimum temperature. |
Record the basin and inlet temperatures for qualified service. Basin Temp detailsShow MoreShow LessThe OEM operator manual provides separate displays for basin temperature and incoming water temperature. Incoming-water measurements are meaningful while water is flowing. Note the affected station, the cycle phase, and which temperature you observed when giving readings to facilities or service. |
| Cancel?Cycle / Prompts | Prompt
The equipment asks the operator to confirm cancellation. |
Confirm only if intended. Follow the OEM termination and reprocessing instructions for an affected scope. |
| Clear Sys Alarm?System / Electronics | Prompt
A service diagnostic is asking for confirmation before clearing a system alarm. |
Record the underlying alarm and refer it to qualified service. Clear Sys Alarm? detailsShow MoreShow LessSTERIS 4002160/C, p. 70. |
| CompletedCycle / Prompts | Status
The display indicates cycle completion. |
Verify the hookups and follow the manufacturer's completion checks. Completed detailsShow MoreShow LessThe STERIS completion procedure describes the indicator and display after the operator records a passing concentration result. The unloading instructions also require checking the hookups. If a connection came off during the cycle, rerun disinfection as directed by the manufacturer before the scope is used. |
| Flash High ErrorSystem / Electronics | Fault
The integrity check for the high-code flash region failed. |
Record the message and contact qualified service. Flash High Error detailsShow MoreShow LessThe appendix identifies the failed check as a cyclic redundancy check (CRC) for the high-code flash region. Flash Low Error identifies the other region. Record the exact display wording for service. The appendix supplies the distinction without identifying a part to replace. STERIS 4002160/C, p. 70. |
| Flash Low ErrorSystem / Electronics | Fault
The integrity check for the low-code flash region failed. |
Record the message and contact qualified service. Flash Low Error detailsShow MoreShow LessThe cyclic redundancy check (CRC) for the low-code flash region did not pass. This wording helps service distinguish it from Flash High Error and from the separately listed RAM and NVRAM messages. The appendix gives a definition and leaves diagnosis to service. STERIS 4002160/C, p. 70. |
| Flo DisabledFlow / Air | Configuration
Liquid-flow monitoring is switched off. |
Have qualified service verify the configuration. Flo Disabled detailsShow MoreShow LessSTERIS 4002160/C, p. 70. |
| Flow 2 Sen ErrFlow / Air | Fault
The SSG chamber-flow sensor registers flow at cycle start, when it should register none. |
Refer the chamber-flow reading to qualified service. Flow 2 Sen Err detailsShow MoreShow LessThe second flow sensor is associated with the SSG chamber. Its startup alarm identifies unexpected indicated flow before the cycle proceeds. No Fluid Fl2 refers to the same part of the system but signals an absence of detected flow at the point it is expected. STERIS 4002160/C, p. 70. |
| Flow Sen ErrAlso search: Flow Sensor ErrorFlow / Air | Fault
The flow sensor registers flow at cycle start, when it should register none. |
Record whether it occurs at startup. Contact support if it persists or returns. Flow Sen Err detailsShow MoreShow LessThe timing matters. This message is generated when the flow sensor shows flow at startup. The earlier Medivators operator manual discusses a sensor that stays in the on state and gives a startup check for that revision. The separate No Fluid Flo alarm concerns absent flow later in the process. STERIS 4002160/C, p. 70. Earlier Medivators documentation: Medivators 50097-027 Rev E, pp. 76–80. |
| HLD FailChemistry / SSG | Status
The recorded disinfectant concentration test did not pass. |
Keep the scope out of clinical use and follow the OEM procedure to rerun disinfection. HLD Fail detailsShow MoreShow LessThe manufacturer’s completion procedure treats a failed minimum-concentration check as unsuccessful processing for the affected scope. It calls for rerunning disinfection. Record the failure and follow the applicable operator and test-strip instructions before returning the endoscope to patient use. |
| HLD PassChemistry / SSG | Status
The disinfectant concentration test was recorded as passing. |
Verify the full OEM completion procedure before releasing the scope. HLD Pass detailsShow MoreShow LessA passing strip check is one part of the completion procedure. The STERIS manual then describes the completion indication and checking the hookup connections when unloading. A hookup that became disconnected during the cycle requires rerunning disinfection under the same manual. |
| HLD Pass or Fail?Chemistry / SSG | Prompt
The unit is waiting for the operator to record the disinfectant concentration test result. |
Follow the applicable test-strip instructions and enter the actual result. HLD Pass or Fail? detailsShow MoreShow LessThe prompt is the point at which the operator records the test result. The STERIS manual describes taking the sample from the designated port and interpreting the strip using its instructions. Enter a passing result only when the test demonstrates that the applicable minimum concentration was met. |
| HLD Sampl ErChemistry / SSG | Fault
A sensor at the disinfectant sampling port reports an incorrect reading. |
Refer the sampling fault to qualified service. |
| Leak DisabledAlso search: Sth DisabledLid / Leak Test | Configuration
The reprocessor's leak-test function is switched off. |
Have qualified service verify the configuration. Leak Disabled detailsShow MoreShow LessSTERIS describes the reprocessor’s built-in leak test as an additional safeguard. The operator manual calls for a separate manual leak test or validated dry leak test before reprocessing. If the disabled indication is unexpected, have qualified service review the unit’s configuration. STERIS 4002160/C, p. 71, p. 41. Earlier Medivators documentation: Medivators 50097-027 Rev E, Medivators 50097-037 Rev E, pp. 129–130. |
| Leak FailAlso search: Sheath FailLid / Leak Test | Fault
The reprocessor detected a failed leak test. |
Follow the manual's major- or minor-leak procedure and the endoscope manufacturer's instructions. Leak Fail detailsShow MoreShow LessThe STERIS manual distinguishes two outcomes. A major leak stops the cycle and the manual calls for resolving the issue before starting a new disinfection cycle. A minor leak can be reported after the cycle has run to completion. The manual still directs the operator to resolve the leak. The in-machine leak test is a backup to pre-reprocessing leak testing. Use the applicable STERIS leak procedure together with the endoscope manufacturer’s instructions to determine how to handle the affected scope. STERIS 4002160/C, p. 71, p. 41, p. 42. Earlier Medivators documentation: Medivators 50097-027 Rev E, Medivators 50097-037 Rev E, pp. 129–130. |
| Leak Sen ErrAlso search: Shth Sen Err, Sheath Sensor ErrorLid / Leak Test | Fault
The leak-test system measured pressure at cycle start. |
Record the message and have qualified service assess the leak-test system. Leak Sen Err detailsShow MoreShow LessThe appendix distinguishes this startup pressure reading from a failed leak test. Older manuals call it Shth Sen Err and discuss pressure already present at startup on units with the optional tester. Record whether the message appears before processing starts and have qualified service verify the applicable procedure. STERIS 4002160/C, p. 71. Earlier Medivators documentation: Medivators 50097-027 Rev E, Medivators 50097-037 Rev E, pp. 129–130. |
| Lid ajarLid / Leak Test | Fault
The system detects an open lid during a cycle. |
When safe, check that the reprocessor lid closes fully and is unobstructed. Lid ajar detailsShow MoreShow LessThe earlier operator manual describes this message during startup and directs trained operators to check whether anything is preventing the lid from closing. The STERIS manual cautions against opening lids during disinfection because of vapor exposure. For a fault mid-cycle, follow the applicable interruption procedure before accessing the basin. STERIS 4002160/C, p. 70, p. 39, p. 42, p. 43. Earlier Medivators documentation: Medivators 50097-027 Rev E, pp. 76–80. |
| Lid DisabledLid / Leak Test | Configuration
Lid-position monitoring is switched off. |
Have qualified service verify the configuration. Lid Disabled detailsShow MoreShow LessSTERIS 4002160/C, p. 70. |
| Low AlcoholChemistry / SSG | Warning
Alcohol supply is low. |
Check the alcohol reservoir and replenish it per the operator manual if used. Low Alcohol detailsShow MoreShow LessThe daily startup section explains how to check and replenish the alcohol reservoir on installations that use it. STERIS also cautions that detergent and alcohol reservoirs should be serviced separately so the correct fluid goes into each container. Follow the reservoir instructions in the revision for the installed unit. |
| Low ChamberWater / Drain | Fault
The basin did not reach the level sensor before the fill timer expired. |
Identify the phase. During rinse, have the water supply and filters checked. During disinfection, consult the applicable chemistry and fill-path checks. Low Chamber detailsShow MoreShow LessStart with the phase shown on the display or recorded in the log. For a rinse-phase event, the earlier Medivators material identifies restricted prefilters or internal filters, inadequate water supply, and insufficient air pressure as possibilities. The STERIS manual describes checking inlet pressure while water is flowing and assessing the pressure drop across prefilters. During disinfection, the earlier manuals also point to possible restriction at the disinfectant screen or debris at a filter connection. An alarm can identify a fill problem without locating the restriction. Tell service whether it occurred during rinse or disinfection and, when safely observable, whether the basin actually filled. STERIS 4002160/C, p. 70, p. 28, p. 32, p. 33. Earlier Medivators documentation: Medivators 50097-027 Rev E, pp. 76–80, Medivators 50097-037 Rev E, pp. 126–128. Related service reading: Low Chamber service article. Use the manufacturer instructions for the installed unit. |
| Low Det.Also search: Low DetergentChemistry / SSG | Warning
Detergent supply is low. |
Check and replenish the detergent reservoir per the operator manual if used. Low Det. detailsShow MoreShow Less |
| No Air FlowFlow / Air | Fault
The system detected no airflow. |
When safe, check that the hookup is connected and free of kinks. Use the applicable manual for air-filter checks. No Air Flow detailsShow MoreShow LessThe earlier Medivators manual starts with the hookup connection and routing, including kinks, and the condition and connection of the air filter. STERIS 4002160/C gives the operator procedure for replacing the air filter. The earlier manual routes compressor problems and inadequate external air pressure to service. STERIS 4002160/C, p. 71, p. 65. Earlier Medivators documentation: Medivators 50097-027 Rev E, pp. 76–80, Medivators 50097-037 Rev E, pp. 126–128. |
| No Fluid Fl2Also search: No Fluid FI2, No Fluid F12, No Flow Fl2, No Fluid Flow FL2Flow / Air | Fault
The SSG chamber-flow sensor did not detect liquid flow. |
Record the exact display text and refer a recurring fault to qualified service. No Fluid Fl2 detailsShow MoreShow LessThe service manual associates the FL2 alarm with chamber flow during disinfectant mixing. It lists a restricted disinfection filter and a restriction at the basin drain screen as possible causes. The operator manual identifies this as an SSG-only message. Use the exact display text when reporting it so that service can distinguish it from the separate No Fluid Flo alarm. STERIS 4002160/C, p. 71. Earlier Medivators documentation: Medivators 50097-037 Rev E, pp. 126–128, pp. 129–130. Related service reading: No Fluid Flow service article. Use the manufacturer instructions for the installed unit. |
| No Fluid FloAlso search: No Fluid FlowFlow / Air | Fault
The system detected no liquid flow. |
Record the phase. When safe, check the applicable scope-to-basin hookup for disconnection or pinching. No Fluid Flo detailsShow MoreShow LessDuring flush or rinse, the earlier Medivators manual points to the scope-to-basin connection, restricted water filters, low incoming water pressure and trapped air in the internal filter housing. The STERIS manual documents checking pressure with water flowing, replacing filters and purging the 0.2-micron filter housing. Use the applicable procedures for the installed revision. During disinfection, the older manual also lists a restricted disinfectant screen, a pinched hookup and a blocked scope channel. A disconnected hookup changes the disposition of the endoscope. The STERIS completion instructions require rerunning disinfection when a hookup disconnected during processing. Refer a suspected blocked scope channel or recurring alarm to qualified service. STERIS 4002160/C, p. 71, p. 39, p. 41, p. 28, p. 32, p. 33, p. 61, p. 63, p. 65, p. 37, p. 38. Earlier Medivators documentation: Medivators 50097-027 Rev E, pp. 76–80, Medivators 50097-037 Rev E, pp. 126–128. Related service reading: No Fluid Flow service article. Use the manufacturer instructions for the installed unit. |
| NVRAM ErrSystem / Electronics | Fault
The controller reports an error in nonvolatile memory. |
Record the alarm and contact qualified service. NVRAM Err detailsShow MoreShow LessSTERIS 4002160/C, p. 71. |
| Power onSystem / Electronics | Status
Power was interrupted during a cycle. |
Review the cycle log and verify the final outcome before releasing the scope. Power on detailsShow MoreShow LessThe 4002160/C interruption procedure says the cycle restarts automatically when power is restored and records the event in the log. Review the affected station’s record and check the final outcome under the OEM completion instructions before releasing the endoscope. |
| Printer does not printSystem / Electronics | Symptom
Output is missing from the optional printer. |
Check the paper supply and applicable print operation in the operator manual. Printer does not print detailsShow MoreShow LessThe operator manual gives separate instructions for requesting log output and loading the paper roll. Check whether the issue is limited to paper output or also affects access to the cycle record. Follow facility traceability procedures if the result for an endoscope cannot be verified. |
| Printer output is faintSystem / Electronics | Symptom
The printed record is hard to read. |
Check the printer instructions for the installed configuration. Printer output is faint detailsShow MoreShow LessThe legacy operator manual covers ribbon printers. Follow 4002160/C p. 66 for its documented paper replacement procedure. Medivators 50097-027 Rev E, p. 77. |
| RAM ErrorSystem / Electronics | Fault
The controller reports a RAM fault. |
Record the alarm and contact qualified service. RAM Error detailsShow MoreShow LessSTERIS 4002160/C, p. 71. |
| Reprocessor does not startSystem / Electronics | Symptom
The unit fails to start or the cycle does not begin. |
Record any displayed message and have facilities or service verify the power supply if the unit has no power. Reprocessor does not start detailsShow MoreShow LessUse the startup checks in the applicable operator manual. Escalate unresolved problems. Medivators 50097-027 Rev E, p. 76. |
| Reprocessor is unresponsiveAlso search: Non responsive machineSystem / Electronics | Symptom
The reprocessor is not responding to controls. |
Record the cycle state and contact qualified service. Reprocessor is unresponsive detailsShow MoreShow LessIf a scope was being processed, confirm cycle disposition under the applicable manufacturer instructions. Medivators 50097-027 Rev E, p. 77. |
| Res T HighAlso search: Reservoir Temperature HighTemperature | Legacy
Legacy high-temperature message. Older operator and service manuals describe the measurement point differently. |
Have qualified service confirm which temperature applies to the installed unit. Res T High detailsShow MoreShow LessIn the older operator appendix, the description points to the reservoir. The older service appendix identifies incoming water instead. The STERIS 4002160/C appendix lists Water T High for incoming water. Confirm the installed model configuration and manual revision before assigning a measurement point to Res T High. Medivators 50097-027 Rev E, pp. 84–85. Medivators 50097-037 Rev E, pp. 129–130. |
| Res T LowAlso search: Reservoir Temperature LowTemperature | Legacy
Legacy low-temperature message. Older operator and service manuals describe the measurement point differently. |
Have qualified service confirm which temperature applies to the installed unit. Res T Low detailsShow MoreShow LessThe older operator appendix describes a low reservoir temperature, while the older service appendix describes low incoming-water temperature. Later STERIS documentation lists Water T Low for incoming water. Ask qualified service to confirm the meaning of the legacy display message on the installed unit. Medivators 50097-027 Rev E, pp. 84–85. Medivators 50097-037 Rev E, pp. 129–130. |
| Reset Alarms?Cycle / Prompts | Prompt
The system is asking whether to cancel an alarm indication. |
Record the original message and follow the OEM instructions for that fault. Reset Alarms? detailsShow MoreShow LessSTERIS 4002160/C, p. 71. |
| SNVRAM ErrSystem / Electronics | Fault
The controller reports an error in short nonvolatile memory. |
Record the alarm and contact qualified service. SNVRAM Err detailsShow MoreShow LessSTERIS 4002160/C, p. 71. |
| SSG A lowChemistry / SSG | Fault
The Part A dosing reservoir failed to fill before timeout. |
Follow the OEM low-chemistry prompt and container replacement procedure. SSG A low detailsShow MoreShow LessThis alarm concerns filling the Part A dosing reservoir. The operator manual directs staff to use the system’s low-chemistry indication to identify when a supply container requires replacement. If the alarm returns after the applicable supply procedure, report the message to service as a fill timeout. |
| SSG A Sen ErChemistry / SSG | Fault
The Part A dosing reservoir sensors report an invalid reading. |
Refer to qualified service. SSG A Sen Er detailsShow MoreShow LessSTERIS 4002160/C, p. 71. |
| SSG B lowChemistry / SSG | Fault
The Part B dosing reservoir failed to fill before timeout. |
Follow the OEM low-chemistry prompt and container replacement procedure. SSG B low detailsShow MoreShow LessThis alarm concerns filling the Part B dosing reservoir. The operator manual directs staff to use the system’s low-chemistry indication to identify when a supply container requires replacement. Report a recurring fill timeout to service, including the exact A or B designation. |
| SSG B Sen ErChemistry / SSG | Fault
The Part B dosing reservoir sensors report an invalid reading. |
Refer to qualified service. SSG B Sen Er detailsShow MoreShow LessSTERIS 4002160/C, p. 71. |
| SSG Cmd ErrChemistry / SSG | Fault
The SSG did not recognize a command from the reprocessor. |
Refer to qualified service. SSG Cmd Err detailsShow MoreShow LessThe manual suggests a possible software issue. STERIS 4002160/C, p. 71. |
| SSG Comm ErrChemistry / SSG | Fault
The SSG detected a communication failure with the reprocessor. |
Refer to qualified service. SSG Comm Err detailsShow MoreShow LessThe direction of detection matters when recording the fault. Here the SSG reports trouble communicating with the reprocessor. With SSG Missing, the reprocessor is the side reporting that it cannot see the SSG. The appendix defines both messages without specifying a field repair. STERIS 4002160/C, p. 71. |
| SSG CRC FailChemistry / SSG | Fault
The SSG failed its software integrity check. |
Refer to qualified service. SSG CRC Fail detailsShow MoreShow LessSTERIS 4002160/C, p. 71. |
| SSG Data ErChemistry / SSG | Fault
SSG data is out of range and does not agree with the reprocessor's data. |
Refer to qualified service. SSG Data Er detailsShow MoreShow LessSTERIS 4002160/C, p. 72. |
| SSG Dose ErrChemistry / SSG | Fault
The reprocessor advanced before the SSG confirmed dosing completion. |
Refer to qualified service. SSG Dose Err detailsShow MoreShow LessThe appendix identifies lost SSG data or a program sequencing error as possible explanations. STERIS 4002160/C, p. 72. |
| SSG Fill ErrChemistry / SSG | Fault
The reprocessor advanced before the SSG finished filling. |
Refer to qualified service. SSG Fill Err detailsShow MoreShow LessSTERIS 4002160/C, p. 72. |
| SSG HW ErrChemistry / SSG | Fault
The SSG detected a hardware fault. |
Refer to qualified service. SSG HW Err detailsShow MoreShow LessThe appendix includes a missing capacitive sensor as one possible hardware condition. STERIS 4002160/C, p. 72. |
| SSG MissingChemistry / SSG | Fault
The reprocessor cannot communicate with the SSG. |
Refer to qualified service. SSG Missing detailsShow MoreShow LessThe reprocessor reports that it cannot communicate with the SSG. In the separate SSG Comm Err message, the SSG reports the communication failure. Tell service which wording appeared and when it occurred. STERIS 4002160/C, p. 72. |
| SSG Rsvr ErrChemistry / SSG | Fault
A dosing-reservoir sensor still reads high after the disinfection cycle. |
Refer to qualified service. SSG Rsvr Err detailsShow MoreShow LessThe appendix describes a sensor reading that remains high after disinfection, when the dosing reservoir should have emptied. It describes the reported state. The message by itself does not establish the actual liquid level or which component needs attention. STERIS 4002160/C, p. 72. |
| Task UnfinChemistry / SSG | Fault
The SSG did not finish filling or dosing. |
Refer to qualified service. Record the phase or operation. Task Unfin detailsShow MoreShow LessSTERIS 4002160/C, p. 72. |
| Time ErrSystem / Electronics | Fault
The controller reports a time-base error. |
Record the message and contact qualified service. Time Err detailsShow MoreShow LessSTERIS 4002160/C, p. 72. |
| Water flows continuously from hookupsAlso search: Continuous water flowFlow / Air | Symptom
The older troubleshooting section includes unintended continuous flow from the hookups. |
Keep the affected station out of use and contact qualified service. Water flows continuously from hookups detailsShow MoreShow LessMedivators 50097-027 Rev E, p. 76. |
| Water T HighTemperature | Fault
Incoming water is above the permitted temperature. |
Have facilities or qualified service verify the inlet temperature with water flowing, using the applicable manual. Water T High detailsShow MoreShow LessThe operator manual directs staff to read incoming-water temperature while water is flowing into the basin. It provides an inlet range for the installed version and displays basin temperature separately. Use the current manual for the unit when checking the inlet reading. Report a recurring high reading to facilities or qualified service. STERIS 4002160/C, p. 72, p. 32, p. 33. Earlier Medivators documentation: Medivators 50097-027 Rev E, Medivators 50097-037 Rev E, pp. 129–130. |
| Water T LowTemperature | Fault
Incoming water is below the permitted temperature. |
Have facilities or qualified service verify the inlet temperature with water flowing, using the applicable manual. Water T Low detailsShow MoreShow LessAfter a long period of inactivity, the STERIS water-check instructions call for running water through the machine to clear cooler water from the supply line. Measure the inlet temperature with water flowing as specified in the manual. Keep the inlet reading separate from the basin reading when troubleshooting. STERIS 4002160/C, p. 72, p. 32, p. 33. Earlier Medivators documentation: Medivators 50097-027 Rev E, Medivators 50097-037 Rev E, pp. 129–130. |
After a failed or interrupted cycle
Follow the applicable STERIS/Medivators instructions for the type of interruption. For a cancelled cycle, failed disinfectant concentration check, or hookup found disconnected during a cycle, 4002160/C directs operators to rerun disinfection for the affected scope. The manual explains how to verify completion and handle major and minor leaks. STERIS 4002160/C, p. 40, p. 41, p. 42, p. 43.
Manufacturer documentation
STERIS DSD-EDGE product literature lists operator manual 4002160/C. Use the revision applicable to the installed unit. For earlier display wording, see the manufacturer-authored Medivators operator manual 50097-027 Rev E and service manual 50097-037 Rev E. SSG means Single Shot Generator, the subsystem that dispenses disinfectant.
Record the device model and serial number, exact message, station, cycle phase and any relevant log entries when contacting support.