1 Micron Filter MF01-0070

1 Micron Water Pre-Filter: Compatible with Medivators Advantage Plus

 

ADV02 is a compatible filter for OEM Part MF01-0070.

Water filters should be replaced at a minimum every three months for the 1.0 micron and 0.4 micron filters and every six months for the 0.1 micron filter. If the water pressure drops below 40 PSI or the difference between the gauges is greater than 5 PSI the corresponding filter needs to be replaced. The following conditions must be met before changing a filter: 1. No endoscope reprocessing cycles are running; 2. No water line disinfection procedure is running.

 

Instructions for installing Medivators Advantage Plus water filters:

1. Relieve the pressure in the filtration system by opening the bleeder valve on the filter system (rotate down).

2. Check the inlet pressure gauge to ensure that the pressure is at zero.

3. Place an empty container under the 1 micron filter housing and disconnect the Quick Connect fitting at the bottom of the housing. Allowing all of the water to empty out.

4. Using a filter bowl wrench, remove the 1 micron filter housing and discard the used filter cartridge.

5. Place a new filter in the housing, ensuring that the O-ring is in place.

6. Thread the housing onto the header and tighten by hand only.

7. Repeat the procedure to replace the 0.4 micron and 0.1 micron filters.

8. To remove the 0.1 micron filter rotate the used filter cartridge counterclockwise, pull downward, and discard.

9. To install a new 0.1 micron filter insert a new filter cartridge into the housing header and rotate clockwise until the tabs lock into place.

10. Return the bleed valve to the closed position (rotate up).

11. Slowly open the main water supply valve and check for any leaks

*The above is adapted from Medivators guides.

 

Medivators® and Advantage Plus® are registered trademarks of Medivators Inc. MedService Repair is not affiliated with Medivators and any and all references to Medivators® on this website are intended to be for informational purposes only. The products on this site have not been manufactured or approved by Medivators®.

Read more

DSD-EDGE Error Codes & Troubleshooting

Medivators DSD-EDGE Error Codes & Troubleshooting Guide

Look up alarms and display messages for the Medivators DSD-EDGE automated endoscope reprocessor. The guide includes all 48 error-message appendix entries in STERIS manual 4002160/C plus 16 related prompts, log entries, legacy messages and symptoms.

Before troubleshooting: Record the exact message, station (A or B) and cycle phase. Keep an affected endoscope out of patient use until the applicable manufacturer instructions confirm successful reprocessing.

DSD-EDGE error messages and troubleshooting reference
Message or symptomWhat it meansTroubleshooting and next steps
A Dose ErrorChemistry / SSG Fault

The Part A dosing step timed out before the full delivery.

Escalate to qualified service.

A Dose Error detailsShow MoreShow Less

The Single Shot Generator (SSG) dispenses disinfectant components. This alarm identifies a timeout during delivery from the Part A dosing reservoir. An SSG A low alarm points to an earlier step, when that reservoir is being filled. The appendix specifies the point of failure but gives no component diagnosis or operator repair for this message.

STERIS 4002160/C, p. 70, p. 75, p. 76.

AbortedCycle / Prompts Status

A processing cycle was cancelled before it finished.

Keep the scope out of clinical use and follow the OEM procedure to rerun disinfection.

Aborted detailsShow MoreShow Less

The STERIS operator manual distinguishes terminating a cycle from pausing one. Termination ends processing for that scope and leaves the station in its safe mode. The manual requires another disinfection cycle for the affected endoscope before it can be released for patient use.

STERIS 4002160/C, p. 70, p. 43.

Air DisabledFlow / Air Configuration

Airflow monitoring is switched off.

Have qualified service verify the configuration.

Air Disabled detailsShow MoreShow Less

Confirm applicability to the installed unit before using the affected station.

STERIS 4002160/C, p. 70.

Attach RestrictorFlow / Air Prompt

The procedure calls for the scope simulator in place of a scope hookup.

Use the specified restrictor only in the applicable manufacturer procedure.

Attach Restrictor detailsShow MoreShow Less

STERIS 4002160/C, p. 68.

B Dose ErrorChemistry / SSG Fault

The Part B dosing step timed out before the full delivery.

Escalate to qualified service.

B Dose Error detailsShow MoreShow Less

This alarm concerns delivery from the Part B dosing reservoir. In the RAPICIDE PA workflow described in 4002160/C, Part B is the buffer. The SSG B low alarm refers to the earlier reservoir-fill step. Record the exact message so service can distinguish an interrupted dose from a low supply indication.

STERIS 4002160/C, p. 70, p. 30, p. 75, p. 76.

Bas DisabledWater / Drain Configuration

Basin-level monitoring is switched off.

Have qualified service verify the configuration.

Bas Disabled detailsShow MoreShow Less
...
Read more